Main page

miniIRIS 1.6.0

Important! You must have OS administrator privileges to upgrade the miniIRIS software.

Program files are saved in the following locations (which required read/write permission):

  1. Main installation folder for miniIRIS software
  2. ...\Windows\System32
  3. ...\Program Files\Common Files\medi.com
  4. Additionally Windows registry is modified.

Upgrading to the new miniIRIS version is a process that can not be undo later, so please read the following changes carefully and accept any inconvenience.

2020-11-27 (Ver. 1.6.0.15) - available after validation on 2020-12-08

  1. CANON EOS SDK - update to the latest SDK version 13.12 and support for Canon 200D MKII, Canon M50 cameras.
  2. CANON EOS SDK - support for Canon EOS 80D.
  3. Adobe Reader - update to the latest version and attempt to fix the printing problem after the last Win10 update
  4. Backup service - set of patches regarding paths and cooperation with the IRIS Service.
  5. miniIRIS Updates - patch for updates when IRISSerwer.exe is running.
  6. Manual and automatic backup service - the operation after sleeping has been improved.
  7. Other minor bugs were fixed.

2019-09-24 (Ver. 1.6.0.14) - available after validation on 2019-09-25

  1. Refresh the camera image - minor bugs were fixed.

2019-09-16 (Ver. 1.6.0.13) - available after validation on 2019-09-18

  1. Foot switch - Support for type A (Win 7. Win 10) and new type B (Win 10, latest updates)
  2. Sinutronic (USB Laryngoscope Camera) MJPEG - support for the new camera.
  3. Other minor bugs were fixed.

2019-06-24 (Ver. 1.6.0.12) - available after validation on 2019-07-30

  1. Automatic saving reports - fixed problem with dual copy of each report.
  2. Automatic cancel of manual backup has been added (after 15 seconds of inactivity).
  3. IRIS Service (medicom service) – issues related to communication with the server have been removed.
  4. Backup module - problem when the sub-model disconnects where resolved and logs were added.
  5. Added checking of the system requirements during miniIRIS application launch.
  6. The installer has been improved (the process of uninstall the application).
  7. CANON EOS SDK - support for Canon EOS 200D.
  8. Game Capture HD60 - support for the new grabber was added.
  9. LinkView 1200 - support for the new grabber was added.
  10. Other minor bugs were fixed.

2018-04-23 (Ver. 1.6.0.11) - available after validation on 2018-05-30

  1. Full backup for database and all data was added.
  2. Archive functionality have been updated and separated (now available in miniIRIS Settings).
  3. CANON EOS SDK - support for Canon EOS 1300D.
  4. Update modules for CANON cameras.
    • Update SDK CANON EOS to version 3.6.30.
  5. Other minor bugs were fixed.

2016-12-16 (Ver. 1.6.0.10)

  1. Support for Canon EOS 760D with Live View was added.
  2. SDK for all Canon cameras was updated to version 3.5.0 (Windows 7/8.1/10).
  3. DICOM - added possibility to configure many types of export (output) to the DICOM file (*.DCM).
  4. DICOM - support for lossless compression was added.
  5. DICOM - common settings for DICOM output (one for the station, and not for each user).
  6. Security - log-in module has been extended to a wide list of configurable rules to define passwords and enhance security. The three main areas are described below:

    a) Password strength:

    • Can be defined a minimum password length.
    • Can be defined the number of uppercase and lowercase characters.
    • Can be defined obligation to use numbers and special characters.

    b) Log-in:

    • Password expiration can be set.
    • Reminder to change the password can be set.
    • Blocking recent passwords can be set.
    • The maximum number of attempts can be set.

    c) Additionally, administrator can set the time after which the user is automatically logged out.

    New functionality is available in the miniIRIS Settings in the "Security" part.

  7. Upgrading the error reporting system (Eurekalog).
  8. XP compatibility mode and run as an administrator is now cleared for the main EXE file during installation. It is no needed any more.
  9. Support for DVI2USB 3.0 grabber (correct resolution according to the specification).
  10. Added support for LGP Stream Engine grabber.
  11. Support for AVerMedi C875 Capture grabber.
  12. Libraries for HD AVerMedia video cards were updated.
  13. Polish manuals have been updated.
  14. miniIRIS Service (windows service) - fixed error when stopping the service, increased safety.
  15. Firebird 2.1.7 x64 is supported.
  16. Installation on Windows 10 is available (set of problems).
  17. Set of changes for improvement application performance.
  18. Other minor bugs were fixed.

2015-04-01

  1. Support for Canon 50D, 70D with live view and Canon 100D has been added.
  2. Canon SDK has been updated for all cameras to version 2.15 (works only for Windows 7 or above).
  3. Error while loading camera id has been removed (now several cameras can work at the same time).
  4. Support for new video grabbers HD of AVerMedia (including USB 3.0 grabbers) has been added.
  5. Loading * .mp4 files from external devices and conversion to the internal format has been added.
  6. Amendment for creating images with data in the format MotionJPEG for example. Colposcope UC-21.
  7. Minor bugs have been removed.

2015-01-18

  1. The logo, name and address of the company has been changed to medi.com Sp. z o.o.
  2. The update process has been improved so that it works properly on Windows 8.7 without turning off the UAC.
  3. Program IRIS Server program has been improves so that it works properly on Windows 8.7 without turning off the UAC.
  4. Software for remote support has been updated.
  5. Problem with line wrapping in the description of the examination(8188) has been removed.
  6. Minor bugs have been removed and instructions have been updated.

2013-08-22

  1. New filters have been added to the preview of the photo.

    New filters are available:

    • Optional creating black and white image - one function to select or lack: the red channel, the green channel, blue channel, red free filter, cobalt filter (as a blue channel), black and white as the average of the components.
    • Optional image filtering - one function to select or lack: Sharpening filter 5x5, 5x5 Gaussian smoothing, median 5x5.
    • Optional negative image.
  2. Logging of history of personal data changes.

    In option to add/edit patient and in option to add/edit personnel information about the history of changes has been added. In the window you can see data of the person who added information and the person who last modified the data. In addition to the name and surname, date and time of add/change is displayed. Visibility of information depends on the new authority ‘History of personal data changes’. By default, only the administrator has access to this bookmark.

  3. Adjustment of ENDO converter service.

do góry